Hai V.H.,Vietnam National University, Ho Chi Minh City | Hung N.Q.,University of Science National University | Khai B.T.,Osaka University
Kerntechnik | Year: 2015

In this article the development of a gamma spectroscopy system is described using a scintillation detector Nal(Tl) of 3 inch x 3 inch. The readout electronic for the spectroscopy is built from the fast analog-digital conversion of Flash Analog-Digital Converter (Flash-ADC) 250 MHz - 8 bits resolution, and the embedded Field-Programmable Gate Array (FPGA) technology. The embedded VHSIC Hardware Description Language (VHDL) code for FPGA is built in such a way that it works as a multi channel analyser (MCA) with 4096 Digital Charge Integration (DCI) channels. A pulse generator with frequency varying from Hz up to 12 kHz is used to evaluate the time response of the system. Two standard radioisotope sources of 133Ba and 152Eu with multi gamma energies ranging from several tens keV to MeVare used to evaluate the linearity and energy resolution of the system. © Carl Hanser Verlag GmbH & Co. KG.