SPA BIT

Kiev, Ukraine
Kiev, Ukraine
SEARCH FILTERS
Time filter
Source Type

Hardcastle N.,University of Wollongong | Hardcastle N.,University of Wisconsin - Madison | Cutajar D.L.,University of Wollongong | Metcalfe P.E.,University of Wollongong | And 5 more authors.
Physics in Medicine and Biology | Year: 2010

Rectal balloons are used in external beam prostate radiotherapy to provide reproducible anatomy and rectal dose reductions. This is an investigation into the combination of a MOSFET radiation detector with a rectal balloon for real-time in vivo rectal wall dosimetry. The MOSFET used in the study is a radiation detector that provides a water equivalent depth of measurement of 70 m. Two MOSFETs were combined in a face-to-face orientation. The reproducibility, sensitivity and angular dependence were measured for the dual MOSFET in a 6 MV photon beam. The dual MOSFET was combined with a rectal balloon and irradiated with hypothetical prostate treatments in a phantom. The anterior rectal wall dose was measured in real time and compared with the planning system calculated dose. The dual MOSFET showed angular dependence within ±2.5% in the azimuth and +2.5%/-4% in the polar axes. When compared with an ion chamber measurement in a phantom, the dual MOSFET agreed within 2.5% for a range of radiation path lengths and incident angles. The dual MOSFET had reproducible sensitivity for fraction sizes of 2-10 Gy. For the hypothetical prostate treatments the measured anterior rectal wall dose was 2.6 and 3.2% lower than the calculated dose for 3DCRT and IMRT plans. This was expected due to limitations of the dose calculation method used at the balloon cavity interface. A dual MOSFET combined with a commercial rectal balloon was shown to provide reproducible measurements of the anterior rectal wall dose in real time. The measured anterior rectal wall dose agreed with the expected dose from the treatment plan for 3DCRT and IMRT plans. The dual MOSFET could be read out in real time during the irradiation, providing the capability for real-time dose monitoring of the rectal wall dose during treatment. © 2010 Institute of Physics and Engineering in Medicine.


Lerch M.L.F.,University of Wollongong | Petasecca M.,University of Wollongong | Cullen A.,University of Wollongong | Hamad A.,University of Wollongong | And 5 more authors.
Radiation Measurements | Year: 2011

Intensive synchrotron X-ray microbeams form an integral part of microbeam radiation therapy (MRT). MRT is a novel radiation medicine modality being developed for inoperable and otherwise untreatable brain tumours. The extremely high dose rate (∼20 kGy/s), laterally fractionated radiation field and steep dose gradients utilized in this therapy make real-time dosimetry a significant challenge. In order for this treatment to advance to the clinical trial stage of development real-time dosimetry systems must be developed. This paper demonstrates the capabilities of a new dosimetry system based on an epitaxial silicon detector. The system combines high spatial resolution and real-time readout and we have measured the lateral dose profile of the MRT radiation field which incorporates 59 X-ray microbeams. All microbeam peaks and valley regions between two microbeams are clearly resolved. The measured detector response at any point is reproducible to within 0.5% after scaling for the known synchrotron storage ring beam current lifetime. The variation of the lateral dose profile at different depths in a PMMA phantom has been measured with the results compared to those from Penelope Monte Carlo simulations. The trend in the measured response with depth agrees with the simulation data (within the experimental variation of the central five microbeams peaks and valleys measured). However the measured peak-to-valley ratio response is a factor of 4.5 ± 0.1 times lower than that expected. The disagreement was further investigated and shown to be contributed to by charge recombination effects at the low bias voltages used. © 2011 Elsevier Ltd. All rights reserved.


Livingstone J.,University of Wollongong | Prokopovich D.A.,Australian Nuclear Science and Technology Organisation | Lerch M.L.F.,University of Wollongong | Petasecca M.,University of Wollongong | And 7 more authors.
IEEE Transactions on Nuclear Science | Year: 2012

Silicon microdosimeters for the characterisation of mixed radiation fields relevant to the space radiation environment have been under continual development at the Centre for Medical Radiation Physics for over a decade. These devices are useful for the prediction of single event upsets in microelectronics and for radiation protection of spacecraft crew. The latest development in silicon microdosimetry is a family of large-area n-SOI microdosimeters for real-time dosimetry in space radiation environments. The response of n-SOI microdosimeters to 2 MeV H and 5.5 MeV He ions has been studied to investigate their charge collection characteristics. The studies have confirmed 100% yield of functioning cells, but have also revealed a charge sharing effect due to diffusion of charge from events occurring outside the sensitive volume and an enhanced energy response due to the collection of charge created beneath the insulating layer. The use of a veto electrode aims to reduce collection of diffused charge. The effectiveness of the veto electrode has been studied via a coincidence analysis using IBIC. It has been shown that suppression of the shared events allows results in a better defined sensitive volume corresponding to the region under the core electrode where the electric field is strongest. © 1963-2012 IEEE.


Wong J.H.D.,University of Wollongong | Wong J.H.D.,University of Malaya | Carolan M.,University of Wollongong | Carolan M.,Wollongong Hospital | And 6 more authors.
Medical Physics | Year: 2010

Purpose: Intensity modulated radiation therapy (IMRT) allows the delivery of escalated radiation dose to tumor while sparing adjacent critical organs. In doing so, IMRT plans tend to incorporate steep dose gradients at interfaces between the target and the organs at risk. Current quality assurance (QA) verification tools such as 2D diode arrays, are limited by their spatial resolution and conventional films are nonreal time. In this article, the authors describe a novel silicon strip detector (CMRP DMG) of high spatial resolution (200 μm) suitable for measuring the high dose gradients in an IMRT delivery. Methods: A full characterization of the detector was performed, including dose per pulse effect, percent depth dose comparison with Farmer ion chamber measurements, stem effect, dose linearity, uniformity, energy response, angular response, and penumbra measurements. They also present the application of the CMRP DMG in the dosimetric verification of a clinical IMRT plan. Results: The detector response changed by 23% for a 390-fold change in the dose per pulse. A correction function is derived to correct for this effect. The strip detector depth dose curve agrees with the Farmer ion chamber within 0.8%. The stem effect was negligible (0.2%). The dose linearity was excellent for the dose range of 3-300 cGy. A uniformity correction method is described to correct for variations in the individual detector pixel responses. The detector showed an over-response relative to tissue dose at lower photon energies with the maximum dose response at 75 kVp nominal photon energy. Penumbra studies using a Varian Clinac 21EX at 1.5 and 10.0 cm depths were measured to be 2.77 and 3.94 mm for the secondary collimators, 3.52 and 5.60 mm for the multileaf collimator rounded leaf ends, respectively. Point doses measured with the strip detector were compared to doses measured with EBT film and doses predicted by the Philips Pinnacle treatment planning system. The differences were 1.1%±1.8% and 1.0%±1.6%, respectively. They demonstrated the high temporal resolution capability of the detector readout system, which will allow one to investigate the temporal dose pattern of IMRT and volumetric modulated arc therapy (VMAT) deliveries. Conclusions: The CMRP silicon strip detector dose magnifying glass interfaced to a TERA ASIC DAQ system has high spatial and temporal resolution. It is a novel and valuable tool for QA in IMRT dose delivery and for VMAT dose delivery. © 2010 American Association of Physicists in Medicine.


Wong J.H.D.,University of Wollongong | Wong J.H.D.,University of Malaya | Fuduli I.,University of Wollongong | Carolan M.,University of Wollongong | And 6 more authors.
Medical Physics | Year: 2012

Purpose: Intensity modulated radiation therapy (IMRT) utilizes the technology of multileaf collimators to deliver highly modulated and complex radiation treatment. Dosimetric verification of the IMRT treatment requires the verification of the delivered dose distribution. Two dimensional ion chamber or diode arrays are gaining popularity as a dosimeter of choice due to their real time feedback compared to film dosimetry. This paper describes the characterization of a novel 2D diode array, which has been named the magic plate (MP). It was designed to function as a 2D transmission detector as well as a planar detector for dose distribution measurements in a solid water phantom for the dosimetric verification of IMRT treatment delivery. Methods: The prototype MP is an 11 × 11 detector array based on thin (50 μm) epitaxial diode technology mounted on a 0.6 mm thick Kapton substrate using a proprietary drop-in technology developed by the Centre for Medical Radiation Physics, University of Wollongong. A full characterization of the detector was performed, including radiation damage study, dose per pulse effect, percent depth dose comparison with CC13 ion chamber and build up characteristics with a parallel plane ion chamber measurements, dose linearity, energy response and angular response. Results: Postirradiated magic plate diodes showed a reproducibility of 2.1%. The MP dose per pulse response decreased at higher dose rates while at lower dose rates the MP appears to be dose rate independent. The depth dose measurement of the MP agrees with ion chamber depth dose measurements to within 0.7 while dose linearity was excellent. MP showed angular response dependency due to the anisotropy of the silicon diode with the maximum variation in angular response of 10.8 at gantry angle 180°. Angular dependence was within 3.5 for the gantry angles ± 75°. The field size dependence of the MP at isocenter agrees with ion chamber measurement to within 1.1. In the beam perturbation study, the surface dose increased by 12.1% for a 30 × 30 cm2 field size at the source to detector distance (SDD) of 80 cm whilst the transmission for the MP was 99. Conclusions: The radiation response of the magic plate was successfully characterized. The array of epitaxial silicon based detectors with drop-in packaging showed properties suitable to be used as a simplified multipurpose and nonperturbing 2D radiation detector for radiation therapy dosimetric verification. © 2012 American Association of Physicists in Medicine.


Wong J.H.D.,University of Wollongong | Wong J.H.D.,University of Malaya | Knittel T.,Prince of Wales Hospital | Downes S.,Prince of Wales Hospital | And 8 more authors.
Medical Physics | Year: 2011

Purpose: Stereotactic radiosurgery/therapy (SRS/SRT) is the use of radiation ablation in place of conventional surgical excision to remove or create fibrous tissue in small target volumes. The target of the SRT/SRS treatment is often located in close proximity to critical organs, hence the requirement of high geometric precision including a tight margin on the planning target volume and a sharp dose fall off. One of the major problems with quality assurance (QA) of SRT/SRS is the availability of suitable detectors with the required spatial resolution. The authors present a novel detector that they refer to as the dose magnifying glass (DMG), which has a high spatial resolution (0.2 mm) and is capable of meeting the stringent requirements of QA and dosimetry in SRS/SRT therapy. Methods: The DMG is an array of 128 phosphor implanted n+ strips on a p -type Si wafer. The sensitive area defined by a single n+ strip is 20×2000 μ m 2. The Si wafer is 375 μm thick. It is mounted on a 0.12 mm thick Kapton substrate. The authors studied the dose per pulse (dpp) and angular response of the detector in a custom-made SRS phantom. The DMG was used to determine the centers of rotation and positioning errors for the linear accelerator's gantry, couch, and collimator rotations. They also used the DMG to measure the profiles and the total scatter factor (Scp) of the SRS cones. Comparisons were made with the EBT2 film and standard Scp values. The DMG was also used for dosimetric verification of a typical SRS treatment with various noncoplanar fields and arc treatments when applied to the phantom. Results: The dose per pulse dependency of the DMG was found to be <5% for a dpp change of 7.5 times. The angular response of the detector was investigated in the azimuthal and polar directions. The maximum polar angular response was 13.8% at the gantry angle of 320°, which may be partly due to the phantom geometry. The maximum azimuthal angular response was 15.3% at gantry angles of 90° and 270°. The angular response at the gantry angle of 180° was 6.3%. A correction function was derived to correct for the angular dependence of the detector, which takes into account the contribution of the azimuthal and polar angular response at different treatment couch positions. The maximum positioning errors due to collimator, gantry, and couch rotation were 0.2±0.1, 0.4±0.1, and 0.4±0.2 mm, respectively. The SRS cone Scp agrees very well with the standard data with an average difference of 1.2±1.1%. Comparison of the relative intensity profiles of the DMG and EBT2 measurements for a simulated SRS treatment shows a maximum difference of 2.5%. Conclusions: The DMG was investigated for dose per pulse and angular dependency. Its application to SRS/SRT delivery verification was demonstrated. The DMG with its high spatial resolution and real time capability allows measurement of dose profiles for cone applicators down to 5 mm in diameter, both accurately and rapidly as required in typical SRS/SRT deliveries. © 2011 American Association of Physicists in Medicine.


PubMed | SPA BIT, University of Wollongong and Eastern Heart Clinic
Type: Journal Article | Journal: Australasian physical & engineering sciences in medicine | Year: 2016

Coronary angiography is a procedure used in the diagnosis and intervention of coronary heart disease. The procedure is often considered one of the highest dose diagnostic procedures in clinical use. Despite this, there is minimal use of dosimeters within angiographic catheterisation laboratories due to challenges resulting from their implementation. The aim of this study was to compare entrance dose delivery across locally commissioned c-arms to assess the need for real-time dosimetry solutions during angiographic procedures. The secondary aim of this study was to establish a calibration method for the MOSkin dosimeter that accurately produces entrance dose values from the clinically sampled beam qualities and energies. The MOSkin is a real-time dosimeter used to measure the skin dose delivered by external radiation beams. The suitability of the MOSkin for measurements in the angiographic catheterisation laboratory was assessed. Measurements were performed using a 303030cm(3) PMMA phantom positioned at the rotational isocenter of the c-arm gantry. The MOSkin calibration factor was established through comparison of the MOSkin response to EBT2 film response. Irradiation of the dosimeters was performed using several clinical beam qualities ranging in energy from 70 to 105kVp. A total of four different interventional c-arm machines were surveyed and compared using the MOSkin dosimeter. The phantom was irradiated from a normal angle of incidence using clinically relevant protocols, field sizes and source to image detector distance values. The MOSkin was observed to be radiotranslucent to the c-arm beam in all clinical environments. The MOSkin response was reproducible to within 2% of the average value across repeated measurements for each beam setting. There were large variations in entrance dose delivery to the phantom between the different c-arm machines with the highest observed cine-acquisition entrance dose rate measuring 326% higher than the lowest measured cine-acquisition entrance dose rate and with the highest measured fluoroscopic entrance dose rate measuring 346% higher than the lowest measured fluoroscopic entrance dose rate. This comparison of entrance dose delivery across local clinical c-arms demonstrated the disparity in entrance dose delivery across catheterisation laboratories and outlined a need for real-time dose monitoring systems for patients during angiographic procedures. Through use of our calibration method, an average MOSkin calibration of 7.37mV/cGy was established. The calibration method allowed entrance dose to be measured across a range of beam energies and beam qualities without the input of the c-arm beam characteristics. This calibration factor was proven to reproduce entrance dose values to within 5% value of the reference dosimeters response, suggesting potential for further studies and utilisation of the dosimeter in this field.


PubMed | SPA BIT, University of Wollongong and Wollongong Hospital
Type: Evaluation Studies | Journal: Medical physics | Year: 2014

Silicon diode arrays are commonly implemented in radiation therapy quality assurance applications as they have a number of advantages including: real time operation (compared to the film) and high spatial resolution, large dynamic range and small size (compared to ionizing chambers). Most diode arrays have detector pitch that is too coarse for routine use in small field applications. The goal of this work is to characterize the two-dimensional monolithic silicon diode array named MagicPlate-512 (MP512) designed for QA in stereotactic body radiation therapy (SBRT) and stereotactic radio surgery (SRS).MP512 is a silicon monolithic detector manufactured on ap-type substrate. An array contains of 512 pixels with size 0.50.5 mm2 and pitch 2 mm with an overall dimension of 5252 mm2. The MP512 monolithic detector is wire bonded on a printed circuit board 0.5 mm thick and covered by a thin layer of raisin to preserve the silicon detector from moisture and chemical contamination and to protect the bonding wires. Characterization of the silicon monolithic diode array response was performed, and included pixels response uniformity, dose linearity, percent depth dose, output factor, and beam profiling for beam sizes relevant to SBRT and SRS and depth dose response in comparison with ionization chamber.MP512 shows a good dose linearity (R2=0.998) and repeatability within 0.2%. The measured depth dose response for field size of 1010 cm2 agreed to within 1.3%, when compared to a CC13 ionization chamber for depths in PMMA up to 30 cm. The output factor of a 6 MV Varian 2100EX medical linac beam measured by MP512 at the isocenter agrees to within 2% when compared to PTW diamond, Scanditronix point EDD-2 diode and MOSkin detectors for field sizes down to 11 cm2. An over response of 4% was observed for square beam size smaller than 1 cm when compared to EBT3 films, while the beam profiles (FWHM) of MP512 match to within 2% the data measured by radiochromic film.The response of the 2D detector array, MP512, has been evaluated. The properties of the array demonstrated suitability for use as in phantom dosimeter for QA in SRS and SBRT. Although MP512 matches film measurements down to 11 cm2 well, it showed a discrepancy of 4% in the determination of output factors of beams smaller than 0.50.5 cm2 due to the field perturbation generated by the large amount of silicon surrounding the central diode. MP512 is highly capable of measuring beam size (FWHM) and has a discrepancy of less than 1.3% when compared to EBT3 film. A reduction in the detector pitch to less than 2 mm would improve the penumbra reconstruction accuracy at the cost readout electronics complexity.


PubMed | SPA BIT, University of Wollongong and Australian Nuclear Science and Technology Organization
Type: | Journal: Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine | Year: 2014

Circular ion-implanted silicon detector of -particles with a large, 5-cm(2), sensitive area has been developed. An advantage of the detector is that the detector surface is easily cleanable with chemicals. The hardened surface of the detector shows no signs of deterioration of the spectroscopic and electrical characteristics upon repeated cleaning. The energy resolution along the diameters of the detector was (1.00.1)% for the 5.486-MeV -particles. Detailed tests of the charge collection efficiency and uniformity of the detector entrance window were also performed with a 5.5-MeV He(2+) microbeam.


PubMed | SPA BIT, Advacam Ltd., University of Wollongong, University of New South Wales and Wollongong Hospital
Type: Journal Article | Journal: Medical physics | Year: 2015

In this work, the edgeless silicon detector technology is investigated, in combination with an innovative packaging solution, to manufacture silicon detectors with negligible angular response. The new diode is also characterized as a dosimeter for radiotherapy with the aim to verify its suitability as a single detector for in vivo dosimetry as well as large area 2D array that does not require angular correction to their response.For the characterisation of the edgeless-drop-in detector technology, a set of samples have been manufactured with different sensitive areas (1 1 and 0.5 0.5 mm(2)) and different thicknesses (0.1 and 0.5 mm) in four different combinations of top and peripheral p-n junction fabricated on p-type and n-type silicon substrates. The diode probes were tested in terms of percentage depth dose (PDD), dose rate, and linearity and compared to ion chambers. Measurements of the output factor have been compared to film. The angular response of the diodes probes has been tested in a cylindrical PMMA phantom, rotated with bidirectional accuracy of 0.25 under 10 10 cm(2) 6 MV Linac photon beam. The radiation hardness has been investigated as well as the effect of radiation damage on the angular and dose rate response of the diode probes when irradiated with photons from a Co-60 gamma source up to dose of 40 kGy.The PDDs measured by the edgeless detectors show an agreement with the data obtained using ion chambers within 2%. The output factor measured with the smallest area edgeless diodes (0.5 0.5 mm(2)-0.1 and 0.5 mm thick) matches EBT3 film to within 2% for square field size from 10 to 0.5 cm side equivalent distance. The dose rate dependence in a dose per pulse range of 0.9 10(-5)-2.7 10(-4) Gy/pulse was less than -7% and +300% for diodes fabricated on p-type and n-type substrates, respectively. The edgeless diodes fabricated on the p-type substrate demonstrated degradation of the response as a function of the irradiation dose within 5%-15%, while diodes on the n-type substrate show a variation of approximately 30% after 40 kGy. The angular response of all probes is minimal (within 2%) but the N on N and P on P configurations show the best performances with an angular dependence of 1.0% between 0 and 180 in the transversal direction. In this configuration, the space charge region of the passive diode extends from the behind and sidewall toward the anode on the top providing beneficial electric field distribution in the peripheral area of the diode. Such performance has also been tested after irradiation by Co-60 up to 40 kGy with no measurable change in angular response.A new edgeless-drop-in silicon diode fabrication and packaging technology has been used to develop detectors that show no significant angular dependence in their response for dosimetry in radiation therapy. From the characterisation of the diodes, proposed in a wide range of different geometries and configurations, the authors recommend the P-on-P detectors in conjunction with drop in packaging technology as the candidate for further development as single diode probe or 2D diode array for dosimetry in radiotherapy.

Loading SPA BIT collaborators
Loading SPA BIT collaborators