Sicel Technologies

Morrisville, NC, United States

Sicel Technologies

Morrisville, NC, United States
SEARCH FILTERS
Time filter
Source Type

Beyer G.P.,Medical Physics Services LLC. | Beyer G.P.,Sicel Technologies | Kry S.F.,University of Houston | Espenhahn E.,Sicel Technologies | And 3 more authors.
Medical Physics | Year: 2011

Purpose: An implantable metal-oxide semiconductor field effect transistors-based dosimeter has recently been developed for the in vivo monitoring of hypofractionated radiotherapy. This DVS-HFT dosimeter is designed for fraction sizes of 340-950 cGy and can also be used for bis in die fraction monitoring. The current work reports on the testing and evaluation of this dosimeter, including both its basic characteristics as well as its performance during simulated clinical treatment plans. Methods: The authors tested the dose rate dependence of this dosimeter (300 MUmin versus 600 MUmin), the treatment time dependence (4 min per treatment versus up to 60 min per treatment), and the dose and energy dependence (6 and 18 MV irradiations of 700-900 cGy per fraction). Additionally, they irradiated the detectors in-phantom with breast and prostate hypofractionated treatments. Results: The detectors showed no significant dose rate, treatment time, energy, or dose dependence. Furthermore, the detectors were found to perform within manufacturer tolerances for all hypofractionated treatments examined, accurately reporting the measured dose (average disagreement of -0.65%). Conclusions: These dosimeters appear well suited for in vivo monitoring of hypofractionated radiotherapy doses, and thereby, have the potential to improve patient care. © 2011 American Association of Physicists in Medicine.


Lu H.-M.,Massachusetts General Hospital | Mann G.,Sicel Technologies | Cascio E.,Harvard University
Medical Physics | Year: 2010

Purpose: In vivo range verification in proton therapy is highly desirable. A recent study suggested that it was feasible to use point dose measurement for in vivo beam range verification in proton therapy, provided that the spread-out Bragg peak dose distribution is delivered in a different and rather unconventional manner. In this work, the authors investigate the possibility of using a commercial implantable dosimeter with wireless reading for this particular application. Methods: The traditional proton treatment technique delivers all the Bragg peaks required for a SOBP field in a single sequence, producing a constant dose plateau across the target volume. As a result, a point dose measurement anywhere in the target volume will produce the same value, thus providing no information regarding the water equivalent path length to the point of measurement. However, the same constant dose distribution can be achieved by splitting the field into a complementary pair of subfields, producing two oppositely "sloped" depth-dose distributions, respectively. The ratio between the two distributions can be a sensitive function of depth and measuring this ratio at a point inside the target volume can provide the water equivalent path length to the dosimeter location. Two types of field splits were used in the experiment, one achieved by the technique of beam current modulation and the other by manipulating the location and width of the beam pulse relative to the range modulator track. Eight MOSFET-based implantable dosimeters at four different depths in a water tank were used to measure the dose ratios for these field pairs. A method was developed to correct the effect of the well-known LET dependence of the MOSFET detectors on the depth-dose distributions using the columnar recombination model. The LET-corrected dose ratios were used to derive the water equivalent path lengths to the dosimeter locations to be compared to physical measurements. Results: The implantable dosimeters measured the dose ratios with a reasonable relative uncertainty of 1%-3% at all depths, except when the ratio itself becomes very small. In total, 55% of the individual measurements reproduced the water equivalent path lengths to the dosimeters within 1 mm. For three dosimeters, the difference was consistently less than 1 mm. Half of the standard deviations over the repeated measurements were equal or less than 1 mm. Conclusions: With a single fitting parameter, the LET-correction method worked remarkably well for the MOSFET detectors. The overall results were very encouraging for a potential method of in vivo beam range verification with millimeter accuracy. This is sufficient accuracy to expand range of clinical applications in which the authors could use the distal fall off of the proton depth dose for tight margins. © 2010 American Association of Physicists in Medicine.


Patent
Sicel Technologies and North Carolina State University | Date: 2010-06-25

Biocompatible sensors configured for implantation include a first body in communication with a plurality of remote sensor bodies to detect physiological parameters in vivo.


Loading Sicel Technologies collaborators
Loading Sicel Technologies collaborators