Moscow, Russia
Moscow, Russia

Time filter

Source Type

Konoplev E.E.,Luch Scientific and Production Association | Stolyarov V.N.,Roentgenprom JSC | Stolyarov I.N.,Roentgenprom JSC
Biomedical Engineering | Year: 2011

The use of a titanium body for manufacture of rotating anodes of diagnostic X-ray tubes is suggested. The results of metallographic analysis of a composite anode made of titanium, molybdenum, and tungsten are discussed. Titanium is demonstrated to be the optimal material of the anode body. Comparative analysis of temperature fields of two anodes made of titanium and molybdenum is presented. © 2011 Springer Science+Business Media, Inc.


Stolyarov I.N.,Roentgenprom JSC
Biomedical Engineering | Year: 2013

Methods for reduction of the working temperature of X-ray tube bearings are discussed. The results of experimental testing of the thermal resistance to anode heat flux provided by most common models of X-ray tube bearings are presented. © 2013 Springer Science+Business Media New York.


Gadetskii V.O.,Roentgenprom JSC | Kabachenko V.V.,Roentgenprom JSC | Kaperusov S.Y.,Roentgenprom JSC | Kostritskii A.V.,Roentgenprom JSC | Khmel'nikov A.V.,Roentgenprom JSC
Biomedical Engineering | Year: 2011

Equipment and software functionality of the PACS system for the mobile digital mammography room are discussed. The equipment and software were produced at Roentgenprom JSC. The system is designed for automation of both diagnostic and screening mammography of the population. The system workflow, details of production, and operating experience are described. © 2010 Springer Science+Business Media, Inc.


Gurzhiev A.S.,National Research Nuclear University MEPhI | Gurzhiev S.N.,Roentgenprom JSC | Miroshnikov E.M.,Roentgenprom JSC | Novikov V.P.,Roentgenprom JSC
Biomedical Engineering | Year: 2011

An experimental comparative study of the efficiency of three methods for measurement of the resolution of digital radiographic systems is described. An algorithm for subpixel angular projection was constructed. This algorithm allows the resolution of X-ray camera and test object to be measured using a narrow slit and sharp edge. Test results obtained using various methods are compared. © 2011 Springer Science+Business Media, Inc.


Stolyarov V.N.,Roentgenprom JSC
Biomedical Engineering | Year: 2012

The impact of different methods of composite anode tungsten layer processing on X-ray tube radiation intensity is discussed. The results of the analysis of tungsten layer surface samples studied by various methods are reported. The results of the X-ray tube radiation intensity measurement are discussed for each sample. © 2012 Springer Science+Business Media, Inc.


Lobanov I.S.,Roentgenprom JSC | Stolyarov V.N.,Roentgenprom JSC | Stolyarov I.N.,Roentgenprom JSC
Biomedical Engineering | Year: 2012

Application of a vibration diagnosis method for analyzing the quality of rotating anode X-ray tubes is described. The experimental results of testing of vibration characteristics of rotating elements of X-ray tubes are presented. © 2012 Springer Science+Business Media, Inc.


Karyugin D.I.,Roentgenprom JSC | Stolyarov V.N.,Roentgenprom JSC | Stolyarov I.N.,Roentgenprom JSC
Biomedical Engineering | Year: 2014

A method for evaluation of the X-ray tube emitter service life is described. The experimental dependence of the emitter impedance on its service life is discussed. Tungsten emitters of different models were compared analytically. © 2014, Springer Science+Business Media New York.


Gurzhiev S.N.,Roentgenprom JSC | Novikov V.P.,Roentgenprom JSC | Sokolov S.N.,Roentgenprom JSC
Biomedical Engineering | Year: 2014

Use of the Fluoro-ProGraph-RP digital fluorograph for tomosynthesis is described. Specific conditions for application are listed, and results of its functions are presented. In particular, calibration of a focal spot of X-ray generator is discussed, as well as elimination of optical distortions and experimental tests for marginal effects in the case of large objects. The tests performed in this work show that the Fluoro-ProGraph-RP fluorograph can be used in tomosynthesis. © 2014, Springer Science+Business Media New York.