Monica Orthopaedic Research Institute

Antwerpen, Belgium

Monica Orthopaedic Research Institute

Antwerpen, Belgium
SEARCH FILTERS
Time filter
Source Type

Van Haver A.,Ghent University | Van Haver A.,Monica Orthopaedic Research Institute | De Roo K.,Ghent University | De Beule M.,Ghent University | And 7 more authors.
Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine | Year: 2014

To investigate the biomechanical effect of skeletal knee joint abnormalities, the authors propose to implant pathologically shaped rapid prototyped implants in cadaver knee specimens. This new method was validated by replacing the native trochlea by a replica implant on four cadaver knees with the aid of cadaver-specific guiding instruments. The accuracy of the guiding instruments was assessed by measuring the rotational errors of the cutting planes (on average 3.01° in extension and 1.18° in external/internal rotation). During a squat and open chain simulation, the patella showed small differences in its articulation with the native trochlea and the replica trochlea, which could partially be explained by the rotational errors of the implants. This study concludes that this method is valid to investigate the effect of knee joint abnormalities with a replica implant as a control condition to account for the influence of material properties and rotational errors of the implant. © IMechE 2014.